Medical and Dental Consultants’ Association of Nigeria
Home - About us - Editorial board - Search - Ahead of print - Current issue - Archives - Submit article - Instructions - Subscribe - Advertise - Contacts - Login 
  Users Online: 4036   Home Print this page Email this page Small font sizeDefault font sizeIncrease font size
 

  Table of Contents 
ORIGINAL ARTICLE
Year : 2013  |  Volume : 16  |  Issue : 2  |  Page : 253-257

Effects of user experience and method in the inflation of endotracheal tube pilot balloon on cuff pressure


Anesthesiology and Reanimation Department, Firat University, Faculty of Medicine, Elazig, Turkey

Date of Acceptance05-Jul-2012
Date of Web Publication6-Apr-2013

Correspondence Address:
A B Ozer
Anesthesiology and Reanimation Department, Firat University, Faculty of Medicine, Elazig
Turkey
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1119-3077.110139

Rights and Permissions
   Abstract 

Context: Endotracheal tube cuff pressure (ETCP) is recommended to be maintained between 20-30 cmH 2 O limits. While insufficient inflation of ETC may cause aspirations, over-inflation of it may lead to damage in tracheal epithelium. Aims: We planned to investigate the effects of user experience and cuff pressure inflation method differences following endotracheal tube cuff pressure and complaints about it.
Patients and Methods: Two hundred and fifty patients planned for general anaesthesia were included in this study. ETC was inflated by users with different experience according to leakage or pilot balloon palpation techniques. ETCPs were measured by manometer at three periods (5 and 60 minutes after endotracheal intubation, and before extubation). Complaints about it were recorded in post anaesthetic care unit and 24 hours postoperatively.
Results: Though we found experience of user had significant effect on the ETCP regulations, we observed inflation methods did not have any effect. However we found ETCP was higher than normal range with experienced users. A correlation was observed between cuff pressure and anaesthesia duration with postoperative complaints.
Conclusions: Our study concluded that the methods used do not have any significant advantage over one another. While ETC inflated at normal pressure increases as user's experience increases, experience alone is not enough in adjusting ETCP. A manometer should be used in routine inflation of ETC instead of conventional methods. CP and anaesthesia duration have correlations with some postoperative complaints.

Keywords: Endotracheal tube cuff pressure, experience of user, inflation methods


How to cite this article:
Ozer A B, Demirel I, Gunduz G, Erhan O L. Effects of user experience and method in the inflation of endotracheal tube pilot balloon on cuff pressure. Niger J Clin Pract 2013;16:253-7

How to cite this URL:
Ozer A B, Demirel I, Gunduz G, Erhan O L. Effects of user experience and method in the inflation of endotracheal tube pilot balloon on cuff pressure. Niger J Clin Pract [serial online] 2013 [cited 2019 Nov 14];16:253-7. Available from: http://www.njcponline.com/text.asp?2013/16/2/253/110139


   Introduction Top


Endotracheal intubation is frequently performed for general anaesthesia and in the intensive care unit. After endotracheal intubation, ETC is inflated via a pilot balloon thereby preventing tracheal aspiration of oropharyngeal secretions as well as leakage of the ventilated volume. [1] Endotracheal tube cuff pressure is recommended to be maintained between 25-30 cmH 2 O limits. [1],[2],[3] Insufficient inflation of ETC may cause tracheal micro-aspirations and especially ventilator-induced pneumonia in patients that undergo long-term mechanical ventilation. [2],[3] On the other hand, pressure reaching 40-45 cmH 2 O may lead to damages ranging from a decrease in capillary blood flow in tracheal epithelium to ischemia. [4],[5],[6],[7]

In order to prevent insufficient or over-inflation of ETC, one of the following methods have been used in practice: palpation of pilot balloon, elimination of the leakage sound, direct measurement of ETC pressure via a manometer or continuous pressure measurement are used in practice. [4],[8],[9],[10]

Therefore, we decided to investigate the effects of user experience and different cuff pressure inflation methods on the endotracheal tube cuff pressure and postoperative complaints associated with it.


   Patients and Methods Top


After obtaining approval form the local ethics committee, we included in the study 250 patients aged between 16-60 years with American Society of Anaesthesiologists physical status classification I-III that were planned for elective surgery in supine position. We excluded patients scheduled for head-neck surgery. In addition, cases with an operation period less than 60 minutes, urgent or difficult intubation, with endotracheal tube inner diameter less than 7 mm or more than 8.5 mm, patients having Mallampati score of III-IV, with high aspiration risk (full stomach, pregnancy), diabetes mellitus, congenital, laryngotracheal and rheumatic diseases were excluded in the study.

Written informed consent was obtained from all patients during preoperative evaluation. Patients were pre-medicated with intramuscular administration of 0.07 mg/kg midazolam and 0.01 mg/kg-45 minutes before the operation, and thereafter were monitored by electrocardiography, non-invasive blood pressure and pulse oximeter. Following anaesthetic induction with 5 mg/kg sodium thiopental and 0.1 mg/kg vecuronium, the trachea was intubated with the appropriate size (ID = 7.0-8.5 mm) of low-pressure high-volume cuffed endotracheal tubes made of polyvinylchloride. Maintenance of anaesthetic was carried out with desflurane in 60% O 2 -air mixture and 2-4 μg/kg fentanyl. Respiration was adjusted as ETCO 2 30-40 mmHg and maintained mechanically. During emergence from anaesthesia, intravenous neostigmine 0.04 mg/kg, and when necessary, 0.01 mg/kg atropine were used to eliminate the residual effect of muscle relaxant.

After intubation, endotracheal tube cuff was inflated by two different methods, namely till when the leakage sound was discontinuated or pilot balloon palpation by users. These users had different experience levels i.e., anaesthesia technicians, anaesthesiologist research assistant doctors and specialist anaesthesiologists. Anaesthesia technicians were divided into two groups based on their work experience, which is less than and more than 2 years. Anaesthesiologist assistants were also divided into two, which is having work experience less and more than 3 years. Patients were randomly divided into 5 groups by envelope method based on the experience of users.

Group T1 (n = 50): Anaesthesia technician with less experience (5 users).

Group T2 (n = 50): Anaesthesia technician with more experience (5 users).

Group A1 (n = 50): Research assistant doctor with less experience (5 users).

Group A2 (n = 50): Research assistant doctor with more experience (5 users).

Group S (n = 50): Specialist doctor (5 users).

Each user from these groups conducted the application on 10 patients, and endotracheal tube cuffs were inflated according to leakage sound in half of the patients (n = 5) and pilot balloon palpation in the other half (n = 5).

Cuff pressures were measured and recorded with aneroid manometer at the end of expiration at the 5 th and 60 th minutes of endotracheal intubation, and immediately before emergence from anaesthesia. This was performed by an anaesthesiologist other than the users and who did not know to which group the patients belonged to. In cases in which the cuff pressure was above 30 cmH 2 O or below 20 cmH 2 O, the pressure was regulated back to normal limits. The duration of anaesthesia was also recorded. In addition, laryngoscopy and intubation conditions, difficult intubation, and number of attempts of laryngoscopy and intubation were also noted. Laryngoscopy and intubation conditions were assessed according to opening of the chin (impossible, difficult, medium, and ease), condition of vocal cords (fully closed, half closed, moving and fully open, not moving) and to reaction to intubation (severe movement, mild movement, mild diaphragmatic movement, no movement). [11] Presence of sore throat, hoarseness of voice, coughing, and dysphagia were recorded at the PACU and at 24 hours postoperatively "present or absent".

Statistical Package for Social Sciences (SPSS) 12.0 software was used for statistical evaluation. The obtained data were taken as mean ± standard deviation. In inter-group comparison, One-Way ANOVA, Tukey HSD, Kruskal Wallis, and Mann Whitney U test were utilized. Paired-Samples T test were used for inner group comparisons and Pearson correlation analysis was used in correlation. P < 0.05 was established as significant.


   Results Top


A significant difference was not observed between the groups in terms of demographic data [Table 1]. When endotracheal tube cuff pressure was compared between the periods, the value measured at the 5 th minute in all groups were found as significantly higher compared to the values measured at the 60 th minute and before the end of the anaesthesia (P < 0.001) [Table 2]. When compared between groups, ETCP measured at the 5 th minute was observed as significantly higher in Group T1 compared to other four groups (P ≤ 0.001). When ETCP measured in Group T2 at the fifth minute was compared with other groups, it was observed as significantly higher than Group A2 and Group S (P < 0.001). In Group A1, ETCP measured at the 5 th minute was established as significantly higher than Group S (P < 0.001). A significant difference was not observed in the comparison of ETCP measured at other periods between the groups. Cuff pressure distribution is given in [Table 3]. About 70% of mean ETCP at the 5 th minute was above 30 cmH 2 O and the pressure was between the 20-30 cmH 2 O intervals in only 24% of the cases.
Table 1: Demographic data of patients


Click here to view
Table 2: Endotracheal cuff pressure values of patients


Click here to view
Table 3: Cuff pressure errors of patients


Click here to view


When laryngoscopy and intubation conditions were compared between groups, the response given to intubation in Group A1 was observed to be significantly higher than other groups (P < 0.001) [Table 4]. A significant difference was not observed between the groups in terms of difficult intubation, difficult laryngoscopy, and the number of intubation attempts.
Table 4: Intubation conditions of patients


Click here to view


The investigation for the presence of sore throat, hoarseness in voice, coughing, and dysphagia, showed Group A1 as having a significant increase compared to Group T1 in which was recorded only dysphagia in the PACU (P < 0.05). At the 24 hours postoperatively, a significant difference was not observed between the groups, in relation to presence of sore throat, hoarseness in voice, coughing and dysphagia [Figure 1].
Figure 1 : Complaints related to endotracheal intubation. *: P < 0.05, when Group A1 was compared with the Group T1

Click here to view


As distinct from other groups, since both significant intubation reaction and dysphagia in PACU were established in Group A1, a correlation analysis was conducted between them. However, a significant correlation was not found. Correlation analysis was conducted between cuff pressure with PACU and postoperative 24 th hour complaints. A significant correlation was observed between the cuff pressure at the 5 th minute and sore throat in the PACU (P = 0.001), hoarseness in voice (P < 0.05), coughing (P < 0.05) and coughing at the postoperative 24 th hour (P < 0.05). A significant correlation was also found between the cuff pressure measured at the 60 th minute and coughing in PACU (P < 0.05) and sore throat in postoperative 24 th hour (P < 0.01) and between cuff pressure before emergence and coughing in PACU (P < 0.05). In addition, a correlation was established between operation duration and coughing in PACU (P = 0.001), sore throat at the postoperative 24 th hour (P = 0.001) and coughing (P < 0.05).

There were no significant effects found on ETCP of the inflation of ETC with leak sound or pilot balloon palpation [Table 3].


   Discussion Top


In order to prevent some undesired effects of endotracheal intubation such as sore throat, aspiration and stenosis, ETC must be inflated and maintained at a certain pressure interval. In our study, the two methods most commonly used in practice for the inflation of ETC were investigated with users having different levels of experience. While significant difference was not observed between the two methods applied by users having the same level of experience, significant decreases were established in cuff pressure values parallel to experience. Although increases occurred in acceptable pressure values in ETCP along with experience, even in specialist physician group erroneous inflation rate was observed as high. In addition, a correlation was found between postoperative complaints with cuff pressure and the duration of anaesthesia.

ETCP average of all groups at the 5 th minute was found as 49 cmH 2 O in our study, and the pressure was within normal values in only 24% of the patients. The remaining 70% of the patients had a pressure of above 30 cmH 2 O. A similar study conducted by Sengupta et al. in three different hospitals in which they applied pilot balloon palpation predominantly for ETC inflation, established ETCP as 20-30 cmH 2 O only in 27% of the patients. [12] They established ETCP as below 20 cmH 2 O in the remaining 23%, above 30 cmH 2 O in 50%, and above 40 cmH 2 O in 27%. Braz et al. measured ETCPs of patients in postoperative care unit (PACU) and intensive care unit (ICU). [13] While cuff pressure had measured as above 40 cmH 2 O in 90.6% of the patients who underwent anaesthesia with nitrous oxide and 45.4% of patients who underwent anaesthesia without nitrous oxide in PACU, this rate for patients in ICU was reported as 54.8%.

In the literature, they have been compared routinely used methods (such as pilot balloon palpation and leakage sound) with manometric method in terms of their effects on the ETCP. [14],[15] A study that compares pilot balloon leakage with leakage sound does not exist in literature. In our study, users applied both methods but a significant difference was not observed between the two methods.

Faris et al., in their study, compared assistant anaesthesiologists with consultant anaesthesiologists, and staff nurses with head nurse in order to establish the role of experience in ETCP inflation. [8] The study reported that a difference was not observed between assistant anaesthesiologist and consultant anaesthesiologists but there was a difference between nurses. We found appropriate inflation rate increased with increasing experience of anaesthesia technicians. While a significant difference was not found between experience assistant anaesthesiologists and specialist anaesthesiologists, we established a significant difference between assistant anaesthesiologists with less experience and specialist anaesthesiologists.

In studies by Wujtewicz et al. conducted with users having different levels of experience in 2002 and in recent years, it was reported that errors do not diminish with the passing years in ETCP inflation and there is a tendency to over-inflation of ETC even among highly experienced users. [16] Although, in our study, we established a difference between experience and ETCP, we found that even the applications of anaesthesiology specialists were erroneous at a rate of 56%, 44% of which were over-inflation and 12% were under-inflation.

In evaluations in the PACU and postoperative 24 th hour, it has been demonstrated that sore throat, [14],[15] coughing, [14] hoarseness in voice [14],[17] and hemoptysis [17] complaints were more common in pilot balloon palpation compared to manometric method. In our study, dysphagia was observed at the PACU in Group A1. Since, in patients belonging to this group, reaction given to intubation was more compared to other groups, a correlation test was carried out, but a significant result was not obtained. A significant correlation was established between postoperative complaints with cuff pressures and the duration of anaesthesia.

In this study, we investigated the effects of user's experience and different cuff inflation methods on the ETCP and postoperative complaints associated with it. We observed effects on the ETCP of both leak sound and pilot balloon palpation methods were similar. The more safely ETCP values were obtained with increasing experience of the users. But, all of the obtained values were not completely accurate and safe. Therefore, it was thought this experience does not suffice alone in adjusting ETCP. However, we found that cuff pressure and operation duration have correlations with some postoperative complaints. There were no obtained for exactly correct and safe ETCP values, no matter which user experience and inflation method. We suggest ETCPs are checked via manometer instead of conventional methods.

 
   References Top

1.Henderson J. Airway management in the adult. In: Miller RD, Eriksson LI, Wiener-Kronish JP, Young WL, editors. Miller's Anesthesia. 7 th ed. Philedelphia: Churchill-Livingstone Elsevier; 2010. p. 1573-610.  Back to cited text no. 1
    
2.Seegobin RD, Van Hasselt GL. Endotracheal cuff pressure and tracheal mucosal blood flow: Endoscopic study of effects of four large volume cuffs. BMJ 1984;288:965-8.  Back to cited text no. 2
    
3.Mehta S. Endotracheal cuff pressure. BMJ 1984;288:1763-4.  Back to cited text no. 3
    
4.Nseir S, Duguet A, Copin MC, De Jonckheere J, Zhang M, Similowski T. Continuous control of endotracheal cuff pressure and tracheal wall damage: A randomized controlled animal study. Crit Care 2007;11:R109.  Back to cited text no. 4
    
5.Nesek-Adam V, Mrsic V, Oberhofer D, Grizelj-Stojcic E, Kosuta D, Rasic Z. Post-intubation long-segment tracheal stenosis of the posterior wall: A case report and review of the literature. J Anesth 2010;24:621-5.  Back to cited text no. 5
    
6.De S, De S. Post intubation tracheal stenosis. Indian J Crit Care Med 2008;12:194-7.  Back to cited text no. 6
[PUBMED]  Medknow Journal  
7.Hameed AA, Mohamed H, Al-Mansoori M. Acquired tracheoesophageal fistula due to high intracuff pressure. Ann Thorac Med 2006;3:23-5.  Back to cited text no. 7
    
8.Faris C, Koury E, Philpott J, Sharma S, Tolley N, Narula A. Estimation of tracheostomy tube cuff pressure by pilot balloon palpation. J Laryngol Otol 2007;121:869-71.  Back to cited text no. 8
    
9.Morris LG, Zoumalan RA, Roccaforte JD, Amin MR. Monitoring Tracheal Tube Cuff Pressures in the Intensive Care Unit: A Comparison of Digital Palpation and Manometry. Ann Otol Rhinol Laryngol 2007;116:639-42.  Back to cited text no. 9
    
10.Sole ML, Aragon D, Bennett M, Johnson RL. Continuous measurement of endotracheal tube cuff pressure: How difficult can it be? AACN Adv Crit Care 2008;19:235-43.  Back to cited text no. 10
    
11.Krieg N, Mazur L, Booij LH, Crul JF. Intubation conditions and reversibility of a new non-depolarizing neuromuscular blocking agent, Org-NC45. Acta Anaesthesiol Scand 1980;24:423-5.  Back to cited text no. 11
    
12.Sengupta P, Sessler DI, Maglinger P, Wells S, Vogt A, Durrani J. Endotracheal tube cuff pressure in three hospitals, and the volume required to produce an appropriate cuff pressure. BMC Anesthesiol 2004;4:8.  Back to cited text no. 12
    
13.Braz JR, Navarro LH, Takata IH, Nascimento Junior P. Endotracheal tube cuff pressure: Need for precise measurement. Sao Paulo Med J 1999;117:243-7.  Back to cited text no. 13
    
14.Jain MK, Tripathi CB. Endotracheal tube cuff pressure monitoring during neurosurgery - Manual vs. automatic method. J Anaesthesiol Clin Pharmacol 2011;27:358-61.  Back to cited text no. 14
[PUBMED]  Medknow Journal  
15.Al-Metwalli RR, Al-Ghamdi AA, Mowafi HA, Sadek S, Abdulshafi M, Mousa WF. Is sealing cuff pressure, easy, reliable and safe technique for endotracheal tube cuff inflation? A comparative study. Saudi J Anaesth 2011;5:185-9.  Back to cited text no. 15
[PUBMED]  Medknow Journal  
16.Wujtewicz MA, Sawicka W, Owczuk R, Dylczyk-Sommer A, Wujtewicz M. Tracheal tube cuff pressure depends on the anaesthesiologist's experience. A follow-up study. Anestezjol Intens Ter 2009;41:205-8.  Back to cited text no. 16
    
17.Liu J, Zhang X, Gong W, Li S, Wang F, Fu S. Correlations between controlled endotracheal tube cuff pressure and postprocedural complications: A multicenter study. Anesth Analg 2010;111:1133-7.  Back to cited text no. 17
    


    Figures

  [Figure 1]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]


This article has been cited by
1 Comparison of tracheal tube cuff pressure with two techniques: fixed volume versus pilot balloon palpation
Farzad Rahmani,Hassan Soleimanpour,Ali Zeynali,Ata Mahmoodpoor,Kavous Shahsavari Nia,Jafar Rahimi Panahi,Sarvin Sanaei,Maryam Soleimanpour,Robab Mehdizadeh Esfanjani
Journal of Cardiovascular and Thoracic Research. 2017; 9(4): 196
[Pubmed] | [DOI]
2 The effect of user experience and inflation technique on endotracheal tube cuff pressure using a feline airway simulator
Donna M. White,José I. Redondo,Alastair R. Mair,Fernando Martinez-Taboada
Veterinary Anaesthesia and Analgesia. 2017;
[Pubmed] | [DOI]
3 Comparison of endotracheal tube cuff pressure values before and after training seminar
Ayça Tuba Dumanli Özcan,Cihan Döger,Abdülkadir But,Isik Kutlu,Semsi Mustafa Aksoy
Journal of Clinical Monitoring and Computing. 2017;
[Pubmed] | [DOI]
4 Evidence Based Use of Cuffed Endotracheal Tubes in Children
Lisa A. Herbinger
Journal of PeriAnesthesia Nursing. 2017;
[Pubmed] | [DOI]
5 Intraoperative management of ETT and LMA cuff pressures: a survey of anaesthetists’ knowledge, attitude and current practice
A de Castro,PD Gopalan
Southern African Journal of Anaesthesia and Analgesia. 2016; 22(5): 151
[Pubmed] | [DOI]
6 Endotracheal Tube Cuff Pressure Assessment: Education May Improve but not Guarantee the Safety of Palpation Technique
Seyed Alireza Seyed Siamdoust,Masood Mohseni,Arash Memarian
Anesthesiology and Pain Medicine. 2015; 5(3)
[Pubmed] | [DOI]
7 Estimation of Endotracheal Tube Cuff Pressure in a Large Teaching Hospital in Ghana
Ebenezer Owusu Darkwa,Frank Boni,Eugenia Lamptey,Yaw Adu-Gyamfi,Christian Owoo,Robert Djagbletey,Alfred Edwin Yawson,Edmund Ayesu,Daniel Akwanfo Yaw Sottie
Open Journal of Anesthesiology. 2015; 05(12): 233
[Pubmed] | [DOI]



 

Top
  
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
   Patients and Methods
   Results
   Discussion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed2459    
    Printed85    
    Emailed1    
    PDF Downloaded568    
    Comments [Add]    
    Cited by others 7    

Recommend this journal