Year : 2013  |  Volume : 16  |  Issue : 3  |  Page : 279-284

Accuracy of impressions with different impression materials in angulated implants


1 Department of Prosthodontics, Navodaya Dental College, Raichur, India
2 Department of Conservative dentistry, Navodaya Dental College, Raichur, India
3 Director, Branemark Osseointegration Center India, M G Road, Secunderabad, India
4 Department of Prosthodontics, S Nijalingappa institute of dental sciences, Gulbarga, India
5 Director, Branemark Osseointegration Center India, Golden Plaza Complex, Court Road, Gulbarga, India

Date of Acceptance18-Sep-2012
Date of Web Publication14-Jun-2013

Correspondence Address:
R Chowdhary
Branemark osseointegration center India, Golden plaza complex, Court Road, Gulbarga
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1119-3077.113447

Rights and Permissions
   Abstract 

Purpose: To evaluate the dimensional accuracy of the resultant (duplicative) casts made from two different impression materials (polyvinyl siloxane and polyether) in parallel and angulated implants.
Materials and Methods: Three definitive master casts (control groups) were fabricated in dental stone with three implants, placed at equi-distance. In first group (control), all three implants were placed parallel to each other and perpendicular to the plane of the cast. In the second and third group (control), all three implants were placed at 10° and 15 o angulation respectively to the long axis of the cast, tilting towards the centre. Impressions were made with polyvinyl siloxane and polyether impression materials in a special tray, using a open tray impression technique from the master casts. These impressions were poured to obtain test casts. Three reference distances were evaluated on each test cast by using a profile projector and compared with control groups to determine the effect of combined interaction of implant angulation and impression materials on the accuracy of implant resultant cast.
Results: Statistical analysis revealed no significant difference in dimensional accuracy of the resultant casts made from two different impression materials (polyvinyl siloxane and polyether) by closed tray impression technique in parallel and angulated implants.
Conclusion: On the basis of the results of this study, the use of both the impression materials i.e., polyether and polyvinyl siloxane impression is recommended for impression making in parallel as well as angulated implants.

Keywords: Angulated implants, implant impression, impression accuracy, impression material


How to cite this article:
Reddy S, Prasad K, Vakil H, Jain A, Chowdhary R. Accuracy of impressions with different impression materials in angulated implants. Niger J Clin Pract 2013;16:279-84

How to cite this URL:
Reddy S, Prasad K, Vakil H, Jain A, Chowdhary R. Accuracy of impressions with different impression materials in angulated implants. Niger J Clin Pract [serial online] 2013 [cited 2019 Nov 18];16:279-84. Available from: http://www.njcponline.com/text.asp?2013/16/3/279/113447


   Introduction Top


In dental implant prosthesis, fabrication of passively fitting prosthesis will lead to the long term success of the restorations. [1] Reproducing the intraoral relationship of implants through impression procedures is the first step in achieving an accurate, passively fitting prosthesis. The critical aspect is to record the 3-dimensional orientation of the implant as it is present intraorally, other than reproducing fine surface detail for successful implant prosthodontics treatment. [2],[3],[4] Magnified dissipation of stresses, due to a lack of passivity has been proposed to be associated with mechanical failure of the restorative components, and of the implants themselves, due to peri-implant bone loss. [5],[6]

Several studies have been documented in literature regarding the accuracy of impression procedures such as open tray pick-up and closed tray transfer techniques, splinting of impression copings, use of different impression materials, angulation and depth of implants, die material accuracy and master cast realization. [1],[7],[8],[9],[10],[11],[12],[13],[14],[15],[16] Two impression techniques, direct and indirect, are currently used to transfer inter-implant dimensions so that the resultant definitive cast duplicates the clinical condition precisely. [17],[18] Research suggests that a direct technique should be used with multiple angulated implants to decrease the distortion. [12] The indirect impression technique can be considered suitable for a parallel or divergent, 2-implant situation. [19] Different impression materials have been examined for use with conventional and implant-supported restorations. [12] Polyether has been the recommended impression material in the past for implant fabrication because of its dimensional stability, rigidity, and tear resistance. [20] Another material frequently used is addition silicone. [21],[22] Addition silicone materials present many of the desirable properties of polyether, and with hydrophilic addition silicones, improved wettability and dimensional stability equal to the characteristics of polyether have been observed. [22],[23] These silicones may be directly comparable to polyether with respect to the quality of impressions for implant fabrication, at a lower cost. However, there are limited and contradictory reports in the peer-reviewed literature comparing these two impression materials. [12],[24],[25],[26]

Techniques to achieve more accurate impressions for patients with multiple dental implants are described in the literature. Most of the research has focused on techniques to improve accuracy with parallel implants. [2],[7],[10],[18],[27],[28],[29] Carr et al. evaluated impression methods for nonparallel implants; however, no research was found comparing techniques for implant impressions at varying degrees of divergence or convergence. [17],[30]

The purpose of this in-vitro study was to analyze the effect of implant angulation and impression material on the accuracy of resultant/duplicative casts by open tray impression technique.


   Materials and Methods Top


The study was conducted to evaluate the dimensional accuracy of the resultant casts made from two different impression materials (polyvinyl siloxane and polyether) by open tray impression technique in parallel and angulated implants.

To determine the effect of combined interaction of implant angulation and impression materials on the accuracy of resultant casts, three master casts were fabricated for three control groups. All three master casts had three implants positioned in a triangular pattern to create a plane. They were numbered as 1, 2 and 3. In first control group, all the three implants were positioned perpendicular to the plane of the cast and parallel to each other at a distance. In the second control group, implants were at an angulation of 10 o to the long axis of the cast. In the third control group, implants were at 15 o angulation to the long axis of the cast.

Type IV dental stone (Kalrock, die stone class IV, Kalabhai Karson Pvt. Ltd. India) was vacuum spatulated using a mechanical spatulator (Bego vacuum mixer, Bego Bremer Goldschlager will GmBA and Co., Bremen, Germany) and poured into a preformed mold and allowed to set to obtain a master cast. Using an engineering radial drilling machine, 3 implant sites were machined in each master cast in a triangular fashion [Figure 1]. In master cast A, the implant sites were placed parallel to each other and perpendicular to the horizontal plane. In master casts B, the implant sites were machined at an angulation of 10° to the long axis of the cast, tilting towards the centre. In master cast C, the implant sites were machined at an angulation of 15° to the long axis of the cast, tilting towards the centre. Three internal hex implant analogs (ADIN dental implants, Israel), 3.75 mm diameter and 10 mm length were fixed in each of the master cast at the previously drilled sites. Three master casts were labeled and prepared for impression procedures [Figure 2].
Figure 1 : Radial drilling machine used for drilling the master cast at specified angulation

Click here to view
Figure 2 : Control groups or master casts with various angulated implants

Click here to view


Three control groups were formed as follows:

Group 1: Parallel placed Implants.
Group 2: Implants at 10° angulations.
Group 3: Implants at 15° angulations.

The distances between the three implant analogs in each of the master cast were measured using a profile projector (Nikon Profile projector Model V-12; Nikon Corp, Tokyo, Japan) by taking the sharp points of the projected hexagonal silhouette as reference points [Figure 3]. All procedures were completed by the same operator.
Figure 3 : Profile projector and image of implant analog as seen on the projector screen

Click here to view


Three inter-analog distances were measured on the master casts.

  1. A1/B1/C1: Inter-implant analog distance between implant analog number 1 and 2.
  2. A2/B2/C2: Inter-implant analog distance between implant analog number 2 and 3.
  3. A3/B3/C3: Inter-implant analog distance between implant analog number 3 and 1.
These distances were recorded as A 1 = 26.900 mm, A 2 = 27.900 mm and A 3 = 28.900 mm in master cast A, B 1 = 26.300 mm, B 2 = 26.150 mm and B 3 = 26.000 mm in master cast B and C 1 = 25.350 mm, C 2 = 25.450 mm and C 3 = 25.250 mm in master cast C respectively and were taken as standard for further comparison.

A total of 30 impressions were made using medium bodied vinyl polysiloxane impression material (Aquasil, Dentsply/ De Trey) and polyether impression material (Impregum Penta, 3M ESPE, Seefeld, Germany). Ten impressions were made for each group (five each with polyvinyl siloxane and polyether impression material respectively). Custom trays were used to make impressions so a total of 30 custom trays were fabricated (10 each for the Master cast A, B and C respectively). [31]

All the impression copings were tightened to the implant analogs on the master casts at 10 Ncm using a torque driver prior to making all the impressions. Before making the impressions, tray adhesive for the polyvinyl siloxane (Universal Tray Adhesive, Zhermack, Italy) and for the polyether (polyether adhesive, 3M ESPE, Germany) was thinly and evenly applied over the inner surface of each tray to extend approximately 2 mm onto the outer surface of the tray along the periphery. The adhesive was allowed to dry before the impressions were made, and the impression was made following the open tray guidelines. [32] Impressions were checked. And thus, a total of fifteen impressions were made with medium-body consistency polyvinyl siloxane impression material and fifteen with the medium body consistency polyether impression material.

Type IV dental stone (Kalrock, die stone class IV, Kalabhai Karson Pvt. Ltd. India) was used to pour these impression and all these resultant casts were stored at room temperature for a minimum of 24 hours before the measurements were made.

A single calibrated examiner blinded to the nature of the impression material and angulation of master cast, examined all the casts to evaluate the positional accuracy of the implant analogs using a profile projector (Nikon Profile projector Model V-12; Nikon Corp, Tokyo, Japan). All the measurements were made under 10× magnifications. The profile projector allowed measurement of linear distances with an accuracy of 0.001 mm (1 um). The distances were measured taking the references of sharp edges of the projected silhouetted form of the analogs. All the three distances were measured for thirty resultant (duplicative) casts and compared with the control group measurements obtained from the three master casts i.e., A 1 , A 2 , A 3 , B 1 , B 2 , B3 and C 1, C 2 , C 3 for respective groups. All the measurements were subjected to

Statistical analysis

Mean and standard deviation were calculated for each inter-analog distance of duplicative casts of all the groups. The considered variables were the angulation of the implants (Parallel and nonparallel) and the type of impression material (Polyvinyl siloxane and Polyether).


   Results Top


The comparison of mean distances (A1, A2 and A3) of duplicative casts obtained from two impression materials (Polyvinyl siloxane and polyether) with those of master cast A (in which implant analogs were placed parallel) shows non-significant difference [Table 1]a,b.
Table 1

Click here to view


The comparison of mean distances (B1, B2 and B3) of duplicative casts obtained from two impression materials (Polyvinyl siloxane and polyether) with those of master cast B (in which implant analogs were placed at 10° angulation) shows non-significant difference [Table 2]a,b.
Table 2

Click here to view


The comparison of mean distances (C1, C2 and C3) of duplicative casts obtained from two impression materials (Polyvinyl siloxane and polyether) with those of master cast C (in which implant analogs were placed at 15° angulation) shows non-significant difference [Table 3]a,b.
Table 3

Click here to view


No significant difference was recorded in the data obtained.


   Discussion Top


One of the major concerns in implant-retained prostheses is with regard to the accuracy of impressions. This plays an essential role in prosthesis-implant adaptation. An accurate working cast for prosthesis components that have optimal adaptation depends on the impression material as well as the transfer techniques. Furthermore, problems related to investing, casting, alloy properties, and clinician skill should not be neglected. In this study, no significant difference in the dimensional accuracy of the resultant (duplicative) casts, which were made using two different impression materials (polyvinyl siloxane and polyether) and in two different conditions of implants placements, one parallel and other angulated implants. [33],[34]

According to Waskewicz et al.[35] and Lorenzoni et al., [36] the original implant position and orientation must be reproduced in the working cast so that best prosthesis fit may be achieved without interfering in the path of prosthesis placement. Clinically divergence or convergence between the implants may be often greater than 8° or 10°. When multiple implants are placed with different angles, the distortion of the impression material on removal may increase. Few studies have reported less accurate impressions with angulated implants than with straight implants using an experimental cast with 4 or 5 implants. [12],[37] Although polyether has been suggested as the material of choice for implant impression procedures, the use of a more elastic impression material, for example a vinyl polysiloxane material, may hypothetically reduce the permanent deformation of impression material determined by the stress between the material and impression copings created when an impression with the copings is removed from internal connection implants. [14]

The distortion of the impression is a concern, in a three dimensional way, in all of the procedures involved in the indirect dental restorations. It can be regarded as absolute or relative, depending on the point of reference from which it is measured, the absolute distortion is considered when the point of reference is external, whereas a relative distortion is measured from a point that is located internal to the system. According to the several studies, in the present investigation, the relative distortion was considered as a study parameter, as the resultant translational distance was measured from one implant analog to another. This kind of measurement can be considered more clinically relevant than the absolute distortion, as implant supported prosthesis usually connects all the abutments to each other.

This study shows that combined interaction of impression material and implant angulation had no effect on the accuracy of the duplicate casts compared to the definitive casts. The results of this study are limited to three implants and may not be relevant for impressions that have higher or lower numbers of implants. Further in-vitro and in-vivo studies are warranted to fully explore the relative merits of one impression materials over the other and to get to a definite conclusion as to which impression material can be the most accurate in different clinical scenarios pertaining to implant impressions.


   Conclusion Top


Polyvinyl siloxane and polyether impression material have similar dimensional accuracy for transfer procedures in parallel and angulated implants. So on the basis of the results of this study, the use of both the impression materials i.e., polyether and polyvinyl siloxane impression is recommended for impression making in parallel as well as angulated implants.

 
   References Top

1.Carr AB, Master J. The accuracy of implant verification casts compared with casts produced from a rigid transfer coping technique. J Prosthodont 1996; 5:248-52.  Back to cited text no. 1
    
2.Barrett MG, de Rijk WG, Burgess JO. The accuracy of six impression techniques for osseointegrated implants. J Prosthodont 1993;2:75-82.  Back to cited text no. 2
    
3.Wee AG. Comparison of impression materials for direct multi-implant impressions. J Prosthet Dent 2000;83:323-31.  Back to cited text no. 3
    
4.Wee AG, Aquilino SA, Schneider RL. Strategies to achieve fit in implant prosthodontics: A review of the literature. Int J Prosthodont 1999;12:167-78.  Back to cited text no. 4
    
5.Sellers G. Direct assembly framework for osseointegrated implant prosthesis. J Prosthet Dent 1989;62:662-8.  Back to cited text no. 5
    
6.Jemt T, Rubenstein JE, Carlsson L, Lang BR. Measuring fit at the implant prosthodontic interface. J Prosthet Dent 1996;75:314-25.  Back to cited text no. 6
    
7.Assif D, Nissan J, Varsano I, Singer A. Accuracy of implant impression splinted techniques: Effect of splinting material. Int J Oral Maxillofac Implants 1999;14:885-8.  Back to cited text no. 7
    
8.Vigolo P, Majzoub Z, Cordioli G. In vitro comparison of master cast accuracy for single-tooth implant replacement. J Prosthet Dent 2000;83:562-6.  Back to cited text no. 8
    
9.Daoudi MF, Setchell DJ, Searson LJ. A laboratory investigation of the accuracy of two impression techniques for single-tooth implants. Int J Prosthodont 2001;14:152-8.  Back to cited text no. 9
    
10.Vigolo P, Majzoub Z, Cordioli G. Evaluation of the accuracy of three techniques used for multiple implant abutment impressions. J Prosthet Dent 2003;89:186- 92.  Back to cited text no. 10
    
11.Akca K, Cehreli MC. Accuracy of 2 impression techniques for ITI implants. Int J Oral Maxillofac Implants 2004;19:517-23.  Back to cited text no. 11
    
12.Assuncao WG, Filho HG, Zaniquelli O. Evaluation of transfer impressions for osseointegrated implants at various angulations. Implant Dent 2004;13:358- 66.  Back to cited text no. 12
    
13.Naconecy MM, Teixeira ER, Shinkai RS, Frasca LC, Cervieri A. Evaluation of the accuracy of 3 transfer techniques for implant-supported prostheses with multiple abutments. Int J Oral Maxillofac Implants 2004;19:192-8.  Back to cited text no. 13
    
14.Vigolo P, Fonzi F, Majzoub Z, Cordioli G. An evaluation of impression techniques for multiple internal connection implant prostheses. J Prosthet Dent 2004;92:470-6.  Back to cited text no. 14
    
15.Vigolo P, Fonzi F, Majzoub Z, Cordioli G. Master cast accuracy in single-tooth implant replacement cases: An in vitro comparison. A technical note. Int J Oral Maxillofac Implants 2005;20:455-60.  Back to cited text no. 15
    
16.Cehreli MC, Akca K. Impression techniques and misfitinduced strains on implant- supported superstructures: An in vitro study. Int J Periodontics Restorative Dent 2006;26:379-85.  Back to cited text no. 16
    
17.Carr AB. Comparison of impression techniques for a two-implant 15° divergent model. Int J Oral Maxillofac Implants 1992;7:468-75.  Back to cited text no. 17
    
18.Assif D, Marshak B, Schmidt A. Accuracy of implant impression techniques. Int J Oral Maxillofac Implants 1996;11:216-22.  Back to cited text no. 18
    
19.Humphries RM, Yaman P, Bloem TJ. The accuracy of implant master casts constructed from transfer impressions. Int J Oral Maxillofac Implants 1990;5:331-6.  Back to cited text no. 19
    
20.Johnson GH, Craig RG. Accuracy of addition silicones as a function of technique. J Prosthet Dent 1986;55:197-203.  Back to cited text no. 20
    
21.Kern M, Rathmer RM, Strub JR. Three dimensional investigation of accuracy of impression materials after disinfection. J Prosthet Dent 1993;70:449-56.  Back to cited text no. 21
    
22.Vassilakos N, Fernandes CP. Surface properties of elastomeric impression materials. J Dent 1993;21:297-301.  Back to cited text no. 22
    
23.Langenwalter EM, Aquilino SA, Turner KA. The dimensional stability of elastomeric impression materials following disinfection. J Prosthet Dent 1990;63:270-6.  Back to cited text no. 23
    
24.Tjan A, Whang SB, Tjan AH, Sarkissian R. Clinically oriented evaluation of the accuracy of commonly used impression materials. J Prosthet Dent 1986;56:4-8.  Back to cited text no. 24
    
25.Pratten DH, Craig RG. Wettability of a hydrophilic addition silicone impression material. J Prosthet Dent 1989;61:197-202.  Back to cited text no. 25
    
26.Panichuttra R, Jones RM, Goodacre C, Mun CA, Moore BK. Hydrophilic poly (vinyl siloxane) impression materials: Dimensional accuracy, wettability, and effect on gypsum hardness. Int J Prosthodont 1991;4:240-8.  Back to cited text no. 26
    
27.Hsu CC, Millstein PL, Stein RS. A comparative analysis of the accuracy of implant transfer techniques. J Prosthet Dent 1993;69:588-93.  Back to cited text no. 27
    
28.Herbst D, Nel JC, Driessen CH, Becker PJ. Evaluation of impression accuracy for osseointegrated implant supported superstructures. J Prosthet Dent 2000;83:555-61.  Back to cited text no. 28
    
29.Spector MR, Donovan TE, Nicholls JI. An evaluation of impression techniques for osseointegrated implants. J Prosthet Dent 1990;63:444-7.  Back to cited text no. 29
    
30.Carr AB. Comparison of impression techniques for a five-implant mandibular model. Int J Oral Maxillofac Implants 1991;6:448-55.  Back to cited text no. 30
    
31.Phillips KM, Nicholls JI, Tsun MA, Rubenstein J. The Accuracy of three implant impression techniques: A three-dimensional analysis. Int J Oral Maxillofac Implants 1994;9:533-40.  Back to cited text no. 31
    
32.Del'Acqua MA, Chávez AM, Amaral AL, Compagnoni MA, Mollo Fde A Jr. Comparison of impression techniques and materials for an implant supported prosthesis. Int J Oral Maxillofac Implants 2010;25:771-6.  Back to cited text no. 32
    
33.Conrad HJ, Pesun IJ, DeLong R, Hodges JS. Accuracy of two impression techniques with angulated implants. J Prosthet Dent 2007;97:349-56.  Back to cited text no. 33
    
34.Choi JH, Lim YJ, Yim SH, Kim CW. Evaluation of the accuracy of implant-level impression techniques for internal-connection implant prostheses in parallel and divergent models. Int J Oral Maxillofac Implants 2007;22:761-8.  Back to cited text no. 34
    
35.Waskewicz GA, Ostrowski JS, Parks VJ. Photoelastic analysis of stress distribution transmitted from a fixed prosthesis attached to osseointegrated implants. Int J Oral Maxillofac Implants 1994;9:405-11.  Back to cited text no. 35
    
36.Lorenzoni M, Pertl C, Penkner K, Polansky R, Sedaj B, Wegscheider WA. Comparison of the transfer precision of three different impression materials in combination with transfer caps for the Frialit-2 system. J Oral Rehabil 2000;27:629-38.  Back to cited text no. 36
    
37.Cabral LM, Guedes CG. Comparative analysis of 4 impression techniques for implants. Implant Dent 2007;16:187-94.  Back to cited text no. 37
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]


This article has been cited by
1 The accuracy of different dental impression techniques for implant-supported dental prostheses: A systematic review and meta-analysis
Tabea Flügge,Wicher Joerd van der Meer,Beatriz Gimenez Gonzalez,Kirstin Vach,Daniel Wismeijer,Ping Wang
Clinical Oral Implants Research. 2018; 29: 374
[Pubmed] | [DOI]
2 Effects of impression levels and trays on the accuracy of impressions taken from angulated implants
Farideh Geramipanah,Majid Sahebi,Maryam Davari,Mohammadreza Hajimahmoudi,Vahid Rakhshan
Clinical Oral Implants Research. 2014; : n/a
[Pubmed] | [DOI]
3 Digital evaluation of the accuracy of impression techniques and materials in angulated implants
Sevcan Kurtulmus-Yilmaz,Oguz Ozan,Tuncer Burak Ozcelik,Ayberk Yagiz
Journal of Dentistry. 2014;
[Pubmed] | [DOI]



 

Top
  
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed3539    
    Printed85    
    Emailed4    
    PDF Downloaded944    
    Comments [Add]    
    Cited by others 3    

Recommend this journal