ORIGINAL ARTICLE
Year : 2016  |  Volume : 19  |  Issue : 2  |  Page : 248-253

Assessment of intraoral image artifacts related to photostimulable phosphor plates in a dentomaxillofacial radiology department


Department of Dentomaxillofacial Radiology, Faculty of Dentistry, Baskent University, Ankara, Turkey

Date of Acceptance21-Apr-2015
Date of Web Publication9-Feb-2016

Correspondence Address:
Dr. A Gulsahi
Department of Dentomaxillofacial Radiology, Faculty of Dentistry, Baskent University, 11. Sokak No. 26 Bahcelievler, Ankara
Turkey
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1119-3077.164338

Rights and Permissions
   Abstract 

Objective: The aim of the present study was to evaluate the presence, frequency, and causes of artifacts in intraoral images obtained using photostimulable phosphor (PSP) plates.
Materials and Methods: A total of 11,443 intraoral images, including 4291 periapical and 7152 bitewing images, acquired over a 6-month period as well as over a month 1-year after the initial imaging were evaluated by a single observer and image artifacts only related to the PSP system were recorded. Before the study, an experienced dentomaxillofacial radiologist and a research assistant assessed a set of image artifacts and agreed on the causes of these artifacts. All unidentified artifacts were reassessed by both researchers before the final decision. The data were analyzed using the statistical software SPSS 11.5.
Results: The total number of images with one or more artifacts was 2344 (20.4%). Of these, 2008 were of adult patients and 336 were of pediatric patients. While movement of the phosphor plate in the disposable pocket was the most common cause of the observed image artifacts in the children, non-uniform image brightness was the most frequently observed artifact in the case of the adults.
Conclusion: The percentage of images with artifacts in the 6th month was lower than that during the 1st month. More significantly, the lowest percentage was obtained 1-year after the initial imaging, owing to the increase in familiarity with the system. Understanding the reasons for the image artifacts and studying ways of preventing are of high clinical importance.

Keywords: Image artifact, intraoral digital imaging, photostimulable phosphor


How to cite this article:
Gulsahi A, Secgin C K. Assessment of intraoral image artifacts related to photostimulable phosphor plates in a dentomaxillofacial radiology department. Niger J Clin Pract 2016;19:248-53

How to cite this URL:
Gulsahi A, Secgin C K. Assessment of intraoral image artifacts related to photostimulable phosphor plates in a dentomaxillofacial radiology department. Niger J Clin Pract [serial online] 2016 [cited 2019 Sep 21];19:248-53. Available from: http://www.njcponline.com/text.asp?2016/19/2/248/164338


   Introduction Top


Digital receptors for intraoral radiography were developed in the late 1980s. All around the world, general dental practitioners are switching from conventional film radiography to digital imaging.[1] The main advantages of digital imaging over conventional radiography are that the former method is faster, allows for real-time imaging and communication, does not require darkroom procedures and chemicals, can avail of image improvement tools, and results in dose reduction.[2],[3],[4],[5],[6],[7],[8],[9] Digital image receptors based on two distinct technologies are used most widely: (1) Solid-state technology (e.g., charge-coupled devices [CCD] or complementary metal oxide semiconductors [CMOS]) and (2) photostimulable phosphor (PSP) technology.[3],[4] In CCD and CMOS systems, a cable usually connects the sensor to the computer, and the image is displayed almost immediately on the computer monitor after the exposure of the sensor.[4]

On the other hand, PSP systems employ phosphor plates that produce a latent image when exposed to radiation. The stored image is transferred to a computer for viewing using a laser scanner. The PSP plates are made of a plastic coated with a polymer base containing embedded phosphor particles. The indirect nature of the image acquisition process in the case of this system results in increased processing times. Further, it necessitates the use of additional equipment.[10] The scanning times range from a few seconds to several minutes, depending on the type of scanner used and the spatial and contrast resolutions of the image.[5] PSP plates are manufactured in a variety of sizes in a manner similar to dental films and are susceptible to bending and scratching during handling.[3],[4],[5],[10],[11] However, PSP plates are intended to be reusable and, therefore, must be handled with greater care than conventional films.[3] The main advantage of PSP systems is that they allow for easy sensor placement within the oral cavity with little discomfort. Therefore, these systems are preferred by dental patients.[10] Image artifacts have been reported in film-based radiography. Digital radiography produces new types of image artifacts, which remain an issue for clinicians.[12] A few examples of the commonly observed image artifacts are listed in [Table 1]. Although a number of studies have investigated the various features of PSP systems.[1],[4],[5],[6],[13],[14],[15] The effects of delayed scanning and visible light on PSP plates,[9],[16],[17],[18] and the effects of damage of PSP plates on image quality,[10],[19] few have assessed PSP image artifacts using representative figures in a clinical setting.[12] Therefore, the aim of the present study was to evaluate the presence, frequency, and causes of PSP artifacts in intraoral images.
Table 1: The most frequent observed image artifact of Digora PSP plate

Click here to view



   Materials and Methods Top


A PSP digital intraoral imaging system (Digora ® Optime, Soredex, Finland) has been used in our university's Department of Dentomaxillofacial Radiology since April 2013. While we had been using digital extraoral imaging since 2002, digital intraoral imaging was a new concept for our faculty. Initially, when using digital intraoral imaging, we adapted the interpacs picture archiving and communication system.

The radiographic and scanning conditions used were those recommended by the respective manufacturers. The PSP system was operated by two experienced radiology technicians and a rotation of students from the department. In the first 2 weeks of implementing PSP imaging, one plate (size 2) was lost, while two plates were damaged during the scanning procedure. Therefore, three new plates had to be used. For infection control, all the intraoral PSP plates were covered with disposable pockets. All the intraoral digital images were obtained using a Prostyle intra X-ray source (65 kVp, 8 mA; Planmeca, Helsinki, Finland). The exposure time was varied, depending on the patient (adult or child) and radiograph type (anterior/posterior or periapical/bitewing). This study was approved by Baskent University Institutional Review Board (Project number D-KA14/01) and supported by University Research Fund.

All the intraoral digital images acquired during a 6-month period (from April, 2013 to October, 2013), including the periapical and bitewing images, as well as those recorded over 1-month 1-year after the initial imaging (April 2014), were evaluated by a single observer (CKS), who is a research assistant in the dentomaxillofacial radiology department. Before the study, an experienced dentomaxillofacial radiologist (AG) and the research assistant assessed a set of image artifacts and agreed on the causes of the observed artifacts. When the cause could not be identified, the possible causes were investigated experimentally. At the end of the evaluation process, all the unidentified artifacts were reassessed by both researchers before the final decision. Only image artifacts related to the PSP system were included in this study.

The data were analyzed using the statistical package of the social sciences (SPSS) version 11.5 for windows (SPSS Inc., Chicago, IL, USA).


   Results Top


A total of 11,443 intraoral images (4291 periapical images and 7152 bitewing images) were evaluated during the initial 6-month period and in the course of a single month 1-year after the initial imaging. Of these images, 9099 (79.6%) had no artifacts, while 2344 (20.4%) had one or more image artifacts. Of all the images evaluated, 10,750 (94%) were obtained from adult patients; of these, 2008 (19%) had one or more image artifacts. Further, 693 images (6%) were obtained from the pediatric patients; of these, 336 (48.4%) had one or more image artifacts. The percentage of image with artifacts was lower in the 6th month than in the 1st month, and it decreased significantly 1-year after the initial imaging (i.e. in April 2014) [Table 2]. During the initial 6-month period, we observed the lowest percentage of artifacts in July. On the other hand, the lowest overall percentage (5%) was observed in April 2014. All the images obtained in July were recorded by an experienced technician because the rotational students did not have classes in this month. The PSP system was completely adopted by April 2014.
Table 2: The percentage of image with or without artifacts according to the months

Click here to view


The types of image artifacts most commonly observed and their number percentages are listed in [Table 3]. These were related to non-uniform image brightness [Figure 1], non-uniform image density [Figure 2], and the movement of the phosphor plates in the disposable packets [Figure 3]. A reduction in the image size displayed after scanning was seen in 0.7% of all the images [Figure 4]. Artifacts related to delayed phosphor plate scanning [Figure 5] and plate surface contamination [Figure 6] were 5th and 6th most frequently observed artifacts. A few artifacts occurred owing to the phosphor plates being exposed from the backside [Figure 7], because of noisy images [Figure 8], the presence of an additional horizontal black line after scanning [Figure 2], and the use of damaged plates [Figure 9]; these were seen in only a few images and were detected only in the 1st month. The artifact caused by the movement of the phosphor plates in the disposable packets was the one observed most frequently in the pediatric age group while related to non-uniform image brightness was the most common one in the adults.
Table 3: The most commonly observed image artifacts and their percentages

Click here to view
Figure 1: Non-uniform image brightness. A bitewing image whose right part is too bright

Click here to view
Figure 2: Non-uniform image density and an image with an additional horizontal black line

Click here to view
Figure 3: Movement of a phosphor plate in a disposable packet

Click here to view
Figure 4: Reduction in the image size (4a). Right (4a) and left (4b) bitewing images of the same patient, respectively; the right one (4a) shows a reduction in size after scanning

Click here to view
Figure 5: Delayed phosphor plate scanning. An image is too bright resulting from a delay in phosphor plate scanning

Click here to view
Figure 6: Plate surface contamination. A right bitewing image whose lower right part has radiopacities resulting from surface contamination with foreign particle

Click here to view
Figure 7: Imaging plate exposed backside. A round metal disk becomes visible on the image due to the backward exposure

Click here to view
Figure 8: A noisy image. Excessive exposure to ambient light between image acquisition and scanning

Click here to view
Figure 9: A physically damaged plate due to scratching

Click here to view


The regions in which the image artifacts were seen are listed in the [Table 4]. The regions in which the artifacts were observed most frequently were the primary maxillary molar, the primary mandibular molar, and the primary maxillary anterior regions. While the artifact related to non-uniform image brightness was seen most commonly in the images of the primary maxillary molars and the primary maxillary anterior regions, that related to the movement of the phosphor plate was seen most frequently in the images of the primary mandibular molar region. Interestingly, the maxillary premolar and mandibular anterior regions were the other regions in which image artifacts were seen frequently, with the artifact related to non-uniform image brightness being the most common in these images.
Table 4: The presence and frequency of the image artifacts according to the region

Click here to view



   Discussion Top


The advent of digital imaging has revolutionized radiology. This revolution is the result of both technologic innovations in the image acquisition processes and the development of networked computing systems for image retrieval and transmission.[3] Despite the development of new advanced methods, intraoral radiography is still the most commonly used the radiographic technique in dental practice.[5] At present, two types of receptors are used for digital intraoral radiography: Solid-state sensors (i.e. CCD or CMOS sensors) that can be used with or without a cord and PSP plates. While sensors based on solid-state technology have been used for more than two decades, systems that use PSP plates have begun to be used in clinical practice only recently. Therefore, the disadvantages associated with PSP plates and the artifacts that occur in the images obtained using them have not been well documented in the literature.

In this study, 2344 (20.4%) of all the 11,443 intraoral images had one or more image artifacts. At the end of the study, it was concluded that the number percentage of the image artifacts was lower in the 6th month than in the 1st month. In addition the percentage decreased significantly 1-year after the initial imaging. In the initial 6-month period, we observed the lowest number of image artifacts in July, because all the images were recorded by an experienced technician in this month. However, the lowest percentage of image artifacts was recorded 1-year later, owing to an increase in familiarity with the system by this time.

While pediatric patients constituted only 6% of the study population, the percentage of images of pediatric patients that exhibited artifacts (48.4%) was higher than the percentage of images of adults that did so (19%). This is probably because of the difficulty in obtaining radiographic images of pediatric patients. However, in digital imaging, it is not essential to retake the radiographs; this protects against excessive exposure to radiation.

While 12.6% of the 11,443 intraoral images recorded in the present study exhibited non-uniform brightness, in another study, 0.37% of the 15,912 scanned PSP images had exhibited errors due to defective plates, and 0.18% had shown scanning-related errors.[12] It is possible that non-uniform image brightness results when some parts of the image are exposed to excessive ambient light as one waits for the other plates to be scanned. Interestingly, all the artifacts were seen on the distal surfaces of the plates, which might have been exposed to ambient light between the cardboard sheaths. Our results showed that non-uniform image density was the second most observed image artifact in the initial 6-month period. This artifact was caused by the partial exposure of the PSP plates to excessive ambient light prior to scanning.[3] Therefore, the plates should not be made to overlap after being exposed. As we were able to determine the case of this artifact, we could prevent it from occurring 1-year after the initial imaging.

While the artifact related to the movement of the phosphor plates in the disposable packets was observed frequently in the present study, other studies have not discussed this type of artifact. It is likely that this artifact is seen only in the images recorded using the Digora ® PSP system because this system requires the use of both cardboard sheaths and disposable plastic envelopes. Although the cardboard sheaths protect the PSP plates, they can cause the movement of the plates, resulting in artifacts. As mentioned previously, this artifact was seen mostly in the images of pediatric patients. This may be because of the difficulty in getting pediatric patients to cooperate fully during radiographic procedures. A decrease in the displayed image size after scanning was also seen in the present study. This was probably because of scanning-related errors, as all the plates were scanned under the same conditions.

Chiu et al.[12] have classified the artifacts observed in intraoral and extraoral images as follows: Those related to operator errors, those related to scanning machine errors, and those related to PSP plate defects. They concluded that artifacts resulting from operator errors were the ones observed most frequently in extraoral images, while those caused by uneven image brightness because of delayed sensor plate scanning were seen most frequently in intraoral images. However, the latter type of artifact was seen in only 0.4% of all the images in the present study. As the PSP plates can tolerate light while being inserted into the scanner, the scanner can be placed in a room exposed to normal daylight or having regular lighting. However, several studies have shown that the higher the light intensity and the longer the exposure, the greater is the loss of information in the plate.[5],[9],[16] It is recommended that PSP plates be scanned no later than 10 min after exposure. If this is not possible, they should be kept in a light-tight environment until scanning.[16] One study showed that plates placed in complete darkness before scanning exhibited no change in image quality, even after several days of storage.[18]

Artifacts caused by factors such as plate surface contamination, backwards phosphor plate exposure, noisy images, the presence of an additional horizontal black line after scanning, and the use of damaged plates were detected in only a few images in the present study. These artifacts occurred owing to the image recorder's lack of familiarity with the PSP system. Indeed, all of these image artifacts could have been eliminated through sufficient training in PSP techniques. In the Digora ® PSP system, phosphor plates exposed backward can be recognized easily because a metal dot is seen in such cases. In contrast to the case with conventional radiography, the resultant image can be corrected by using the mirror function of the image acquisition software that accompanies the PSP system.[3],[12] Further, the sensors can be physically damaged if they are dropped on the floor or if a patient bites hard into their surface.[5] Disposable packets are used for infection control; however, they also act as a barrier against ambient light, and do not provide adequate protection against damage to the plates from bending, pressure from the positioning device, or tooth marks.[10],[12] In this study, only a few PSP plates were damaged and replaced, mostly in the initial stage of using the PSP system. However, a previous study showed that a large number of PSP plates are usually damaged during imaging.[12] Another study concluded that the main reason for the replacement of plates was damage to the phosphor layer.[10]


   Conclusion Top


The change from conventional to digital radiography has been a revolution, and the use of digital imaging systems has increased significantly over the past decades. According to the results of this study, 20.4% of all intraoral images had one or more image artifacts, and the lowest percentage (5%) was obtained 1-year after the initial imaging period. The most frequently observed artifacts were non-uniform image brightness, non-uniform image density, and those caused by movement of the phosphor plates in disposable packets. Understanding the causes of image artifacts and studying methods of preventing these artifacts are of great clinical importance.


   Acknowledgments Top


No external funding, apart from the support of the authors' institution was available for the study.

Authors declare that there are no conflicts of interest in the study.

This article does not contain any studies with animal subjects performed by the any of the authors.

 
   References Top

1.
Wenzel A. A review of dentists' use of digital radiography and caries diagnosis with digital systems. Dentomaxillofac Radiol 2006;35:307-14.  Back to cited text no. 1
    
2.
Malleshi SN, Mahima VG, Raina A, Patil K. A subjective assessment of perceived clarity of indirect digital images and processed digital images with conventional intra-oral periapical radiographs. J Clin Diagn Res 2013;7:1793-6.  Back to cited text no. 2
    
3.
Ludlow JB, Mol A. Digital imaging. In: White SC, Pharoah MJ, editors. Oral Radiology: Principles and Interpretation. St. Louis, Missouri: Mosby, Elsevier; 2014. p. 41-62.  Back to cited text no. 3
    
4.
Berkhout WE, Sanderink GC, Van der Stelt PF. A comparison of digital and film radiography in Dutch dental practices assessed by questionnaire. Dentomaxillofac Radiol 2002;31:93-9.  Back to cited text no. 4
    
5.
Wenzel A, Møystad A. Work flow with digital intraoral radiography: A systematic review. Acta Odontol Scand 2010;68:106-14.  Back to cited text no. 5
    
6.
Brennan J. An introduction to digital radiography in dentistry. J Orthod 2002;29:66-9.  Back to cited text no. 6
    
7.
Van der Stelt PF. Better imaging: The advantages of digital radiography. J Am Dent Assoc 2008;139 Suppl: 7S-13.  Back to cited text no. 7
    
8.
Udupa H, Mah P, Dove SP, McDavid WD. Evaluation of image quality parameters of representative intraoral digital radiographic systems. Oral Surg Oral Med Oral Pathol Oral Radiol 2013;116:774-83.  Back to cited text no. 8
    
9.
Ramamurthy R, Canning CF, Scheetz JP, Farman AG. Impact of ambient lighting intensity and duration on the signal-to-noise ratio of images from photostimulable phosphor plates processed using DenOptix and ScanX systems. Dentomaxillofac Radiol 2004;33:307-11.  Back to cited text no. 9
    
10.
Roberts MW, Mol A. Clinical techniques to reduce sensor plate damage in PSP digital radiography. J Dent Child (Chic) 2004;71:169-70.  Back to cited text no. 10
    
11.
Kitagawa H, Farman AG, Scheetz JP, Brown WP, Lewis J, Benefiel M, et al. Comparison of three intra-oral storage phosphor systems using subjective image quality. Dentomaxillofac Radiol 2000;29:272-6.  Back to cited text no. 11
    
12.
Chiu HL, Lin SH, Chen CH, Wang WC, Chen JY, Chen YK, et al. Analysis of photostimulable phosphor plate image artifacts in an oral and maxillofacial radiology department. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008;106:749-56.  Back to cited text no. 12
    
13.
Stamatakis HC, Welander U, McDavid WD. Physical properties of a photostimulable phosphor system for intra-oral radiography. Dentomaxillofac Radiol 2000;29:28-34.  Back to cited text no. 13
    
14.
Borg E, Attaelmanan A, Gröndahl HG. Image plate systems differ in physical performance. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2000;89:118-24.  Back to cited text no. 14
    
15.
Petrikowski CG. Introducing digital radiography in the dental office: An overview. J Can Dent Assoc 2005;71:651.  Back to cited text no. 15
    
16.
Akdeniz BG, Gröndahl HG, Kose T. Effect of delayed scanning of storage phosphor plates. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2005;99:603-7.  Back to cited text no. 16
    
17.
Martins MG, Neto FH, Whaites EJ. Analysis of digital images acquired using different phosphor storage plates (PSPs) subjected to varying reading times and storage conditions. Dentomaxillofac Radiol 2003;32:186-90.  Back to cited text no. 17
    
18.
Ang DB, Angelopoulos C, Katz JO. How does signal fade on photo-stimulable storage phosphor imaging plates when scanned with a delay and what is the effect on image quality? Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2006;102:673-9.  Back to cited text no. 18
    
19.
Kalathingal SM, Shrout MK, Comer C, Brady C. Rating the extent of surface scratches on photostimulable storage phosphor plates in a dental school environment. Dentomaxillofac Radiol 2010;39:179-83.  Back to cited text no. 19
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7], [Figure 8], [Figure 9]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]


This article has been cited by
1 Definition, classification and retrospective analysis of photostimulable phosphor image artefacts and errors in intraoral dental radiography
Armagan «aliskan,A Pinar Sumer
Dentomaxillofacial Radiology. 2017; : 20160188
[Pubmed] | [DOI]



 

Top
  
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
   Acknowledgments
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed2163    
    Printed51    
    Emailed1    
    PDF Downloaded271    
    Comments [Add]    
    Cited by others 1    

Recommend this journal